I found Ca EXAFS files on my data directory which go out to 600eV above the edge, done using Si111. What's the symptom of the failure with InSb? Could harmonics be the problem? If there's something mechanical going on with the mono such that you go through the max of the rocking curve, then as you go through that max, you get more harmonics, which cause normalization failure. mam On 5/11/2015 11:06 AM, Lisa Bovenkamp wrote:
Hello.
No, the resolution is fine. I can do XANES fine with InSb111. We use those for years. Just had problems with EXAFS because for this long range the flux has to be more stable from the crystals.
Lisa _______________________________________________ Ifeffit mailing list Ifeffit@millenia.cars.aps.anl.gov http://millenia.cars.aps.anl.gov/mailman/listinfo/ifeffit