Hi Weronika,

I think that the problem is due to the fact that the data file stores data in keV so that the small increments in energy at the pre-edge look like very, very small changes if the energy is assumed to be in eV.    So, I think you may just need to specify that the data has energy units of keV rather than eV.

When I read your Fe_Kedge_falconX.... file into XAS Viewer, I also had to tell it that it was XAS data in keV, but then it read in just fine.  If I zoom in on the pre-edge peaks it shows:

  
image.png


I think that agrees with what you see.
 

On Mon, Jun 27, 2022 at 8:40 PM Ofierska Weronika Lucja <weronika.ofierska@erdw.ethz.ch> wrote:

Dear all, 

I am a new user of Athena and encountered some issues with data plotting. 

While plotting the uXAS data for Fe K pre-edge the quality of plotting is somehow less sharp compared to the actual data I plotted using other graphical tool (see below) 


vs what I get from Athena 


As you can see the resolution of the pre-edge in Athena is so poor I cannot resolve the Fe2+/Fe3+ ratio as the peak for Fe3+ is not visible. (I plotted Bruce's data from lepidocrocite to see if it's a common problem and it seems that his data look much better than mine). During my time on synchrotron we actually spent a lot of time to resolve best resolution on the pre edge, so it's not a acquisition problem. 


Does anyone have any idea how to fix this? 

I am attaching the project file and txt file (plot FluoRT2 column) with the original data. 


Thank you for your help, 

Weronika Ofierska 

_______________________________________________
Ifeffit mailing list
Ifeffit@millenia.cars.aps.anl.gov
http://millenia.cars.aps.anl.gov/mailman/listinfo/ifeffit
Unsubscribe: http://millenia.cars.aps.anl.gov/mailman/options/ifeffit


--
--Matt Newville <newville at cars.uchicago.edu> 630-327-7411